Showing 13–24 of 32 results

HbA1c FIA Test

  Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 4%–15%
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of HbA1c reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of HbA1c reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • HbA1c control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of HbA1c reference materials at 4%–15% was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F6100

HbA1c FIA Test

25/kit

Serum, Plasma

Immunofluorescence Analyzer

ACCURACY:
  • Less than 3% intra-run CV
  • Eliminates subjective interpretation by operator
  • Results comparable to lab-based CLIA analyzer
EASE-OF-USE:
  • System uses an SD card to ensure the proper test is being run
  • Able to run whole blood, serum, plasma and urine specimens
  • Touch screen interface for easy operation
  • Lightweight (2.9 kg) and portable
DATA MANAGEMENT:
  • Results presented on screen or via the built-in printer
  • Storage capacity of 10,000 test results
  • SPECIFICATIONS
    Human Machine Interface: 4.3-inch LCD touch screen
    Dimensions: 280(L)×256(W)×102(H)mm
    Weight: 2.9 kg
    Power Adapter: 100-240 VAC, 50-60Hz
    Data output: On screen/Printer
    OPERATION ENVIRONMENT
    Temperature: 10~30℃
    Humidity: ≤70%
    Atmosphere pressure: 85.0~106.0 Kpa
    Operation Altitude: <2000m
    Location: Dry, clean, flat, horizontal surface, away from direct sunlight, mechanical vibration, and strong electromagnetic interference.
    STORAGE ENVIRONMENTAL
    Temperature: 0~50℃
    Humidity: ≤93%
    Storage Elevation: 0~2000m
    OPTICAL DESCRIPTION
    Excitation Light Source: LED 365nm
    Detector: Silicon photodiode

Immunofluorescence Incubator

ACCURACY: • Incubation system control temperature variation ≤± 0.1℃ EASE-OF-USE: • Simple temperature adjustment • Easy incubation time-setting for each channel • Portable and lightweight (1.4 kg) • Incubation completion reminder chime for user convenience • 6-channel incubator to increase throughput
SPECIFICATIONS
Dimensions: 255(L)×135(W)×40(H)mm
Weight: 1.4 kg
Power Adapter: 100-240 VAC, 50 Hz, 60W
Temperature Setting: 20~40℃
Time Setting: 3~30 min
OPERATION CONDITIONS
Temperature: 10~30℃
Humidity: ≤95%
Atmosphere Pressure: 85.0~106.0KPa
Operation Altitude: <2000m
STORAGE CONDITIONS
Temperature: 0~55℃
Humidity: ≤80%
 

LH FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-200.0 mIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of LH reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of LH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • LH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of LH reference materials at 1.0-200.0 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1090

LH FIA Test

25/kit

Serum, Plasma
 

MAU FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 2.0-300.0 mg/L
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of MAU reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of MAU reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  • MAU control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of MAU reference materials at 2.0-300.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F2020

MAU FIA Test

25/kit

Urine

MYO FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 2.5-400.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of MYO reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of MYO reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • MYO control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of MYO reference materials at 20.0-400.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3030

MYO FIA Test

25/kit

Serum, Plasma
 

NT-proBNP FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 15–35000 pg/mL
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of NT-proBNP reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of NT-proBNP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • NT-proBNP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of NT-proBNP reference materials at 15–35000 pg/mL was tested, and the correlation coefficient (R) is ≥0.9900.
REAGENTS AND MATERIALS PROVIDED:
  1. Individually sealed foil pouches containing: a. One test device b. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

PCT FIA Test

Features: PRODUCT FEATURES
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–100.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PCT reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of PCT reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PCT control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PCT reference materials at 0.1–100.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1.  Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0400

PCT FIA Test

25/kit

Serum, Plasma, Whole blood

PRL FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 5.0-4000.0 μIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PRL reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of PRL reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PRL control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PRL reference materials at 5.0-4000.0 μIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1080

PRL FIA Test

25/kit

Serum, Plasma
 

Progesterone FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.30-80.00 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of PROG reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of PROG reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
PROG control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of PROG reference materials at 3.00-80.00 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
Crossed Factors Concentration Test Result
Estradiol 500 ng/mL < 0.3 ng/mL
Testosterone 500 ng/mL < 0.3 ng/mL
Cortisol 500 ng/mL < 0.3 ng/mL
Aldosterone 500 ng/mL < 0.3 ng/mL
Estriol 100 ng/mL < 0.3 ng/mL
17-β Estradiol 100 ng/mL < 0.3 ng/mL
Oestrone 100 ng/mL < 0.3 ng/mL
17-α hydroxyl progesterone 100 ng/mL < 0.3 ng/mL
Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

SNIBE Maglumi 600

Features:
  • Principle: Chemiluminescence Immunoassay (CLIA) System
  • Throughput: up to 180 tests per hour
  • Sample Position: up to 16 sample tubes
  • Continuous sample loading, STAT available
  • Reagent Position: 4 reagents on board
  • Refrigerated reagent area
  • Reagent RFID Tag with built-in master curve and information
  • Calibration stability: maximum of 4 weeks
  • Probe: Liquid Level Detection, Clot Detection

SNIBE Maglumi 800

Features:
  • Principle: Chemiluminescence Immunoassay (CLIA) System
  • Throughput: up to 180 tests per hour
  • Sample Position: up to 40 sample tubes
  • Continuous sample loading, STAT available
  • Reagent Position: 9 reagents on board
  • Refrigerated reagent area
  • Reagent RFID Tag with built-in master curve and information
  • Calibration stability: maximum of 4 weeks
  • Probe: Liquid Level Detection, Clot Detection