AFP FIA Test
Features:
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 3.0-2000.0 ng/mL
- Intra-lot precision was determined by testing of AFP reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot precision was determined by testing of AFP reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
- AFP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of AFP controls at 3.0-2000.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
- Reagents and Materials provided:
- Individually sealed foil pouches containing:
- One test device
- One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
- Individually sealed foil pouches containing:
AMH FIA Test
Features:
-
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Range
- Precision
- Accuracy
- Linearity
- Reagents and Materials provided:
- Individually sealed foil pouches containing:
- One test device
- One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
- Individually sealed foil pouches containing:
CEA FIA Test
Features:
-
-
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Range
- Working range: 0.5-500.0 ng/mL
- Intra-lot precision was determined by testing of CEA reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot precision was determined by testing of CEA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
- CEA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of CEA reference materials at 1-500.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
- There is no Hook effect when the level is 100000.0 ng/mL.
- Individually sealed foil pouches containing:
- One test device
- One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
-
CK-MB FIA Test
Features:
-
-
-
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 1.0-80.0 ng/mL
- Intra-lot Precision Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot Precision Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
- CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of CK-MB reference materials at 2.5-80.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
- REAGENTS AND MATERIALS PROVIDED:
- Individually sealed foil pouches containing: a. One test device b. One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
-
-
CRP FIA Test
Features:
-
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 2.5–200 mg/L
- Intra-lot Precision Intra-lot precision was determined by testing of CRP reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot Precision Inter-lot precision was determined by testing of CRP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
- CRP control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of CRP reference materials at 2.5–200 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
- REAGENTS AND MATERIALS PROVIDED:
- Individually sealed foil pouches containing: a. One test device b. One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
CATALOG # | PRODUCT | QUANTITY | SPECIMEN |
F0410 |
CRP FIA Test |
25/kit |
Serum, Plasma, Whole blood |
cTnI FIA Test
Features:
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 0.1–50.00 ng/mL
- Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
- cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
-
- Individually sealed foil pouches containing: a. One test device b. One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
CATALOG # | PRODUCT | QUANTITY | SPECIMEN |
F3002 |
cTnI FIA Test |
25/kit |
Plasma, Whole blood |
Cys-C FIA Test
Features:
-
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 0.2-8.0 mg/L
- Intra-lot precision was determined by testing of Cys-C reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot precision was determined by testing of Cys-C reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
- Cys-C control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of Cys-C reference materials at 0.4-8.0 mg/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
-
- Individually sealed foil pouches containing:
- One test device
- One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
- Individually sealed foil pouches containing:
D-Dimer FIA Test
Features:
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 0.1–10.0 mg/L
- Intra-lot Precision Intra-lot precision was determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot Precision Inter-lot precision was determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
- D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
- REAGENTS AND MATERIALS PROVIDED:
- Individually sealed foil pouches containing: a. One test device b. One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
CATALOG # | PRODUCT | QUANTITY | SPECIMEN |
F3010 |
D-Dimer FIA Test |
25/kit |
Serum, Plasma |
Ferritin FIA Test
Features:
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 1.0-1000.0 ng/mL
- Intra-lot Precision Intra-lot precision was determined by testing of Ferritin reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot Precision Inter-lot precision was determined by testing of Ferritin reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
- Ferritin control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of Ferritin reference materials at 2.0-1000.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
- REAGENTS AND MATERIALS PROVIDED:
- Individually sealed foil pouches containing: a. One test device b. One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
CATALOG # | PRODUCT | QUANTITY | SPECIMEN |
F5010 |
Ferritin FIA Test |
25/kit |
Serum, Plasma |
fPSA FIA Test
Features:
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 0.1-30.0 ng/mL
- Intra-lot precision was determined by testing of fPSA reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot precision was determined by testing of fPSA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
- fPSA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
- A serial concentration of fPSA reference materials at 0.1-30.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
- Reagents and Materials provided:
- Individually sealed foil pouches containing:
- One test device
- One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
- Individually sealed foil pouches containing:
FSH FIA Test
Features:
-
- Quantitative, non-subjective results
- Increased sensitivity and broader detection range compared to rapid tests
- Working range: 0.30-200.00 mIU/mL
- Intra-lot Precision Intra-lot precision was determined by testing of FSH reference materials using 10 test devices from the same lot. CV ≤ 15%.
- Inter-lot Precision Inter-lot precision was determined by testing of FSH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
- FSH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%
- A serial concentration of FSH reference materials at 1.00-200.00 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
- REAGENTS AND MATERIALS PROVIDED:
- Individually sealed foil pouches containing: a. One test device b. One desiccant
- Detection buffer tubes
- SD card
- Instructions for Use
- Clock or timer
- RaFIA Immunofluorescence Analyzer
- RaFIA Immunofluorescence Incubator
CATALOG # | PRODUCT | QUANTITY | SPECIMEN |
F1100 |
FSH FIA Test |
25/kit |
Serum, Plasma |