AFP FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
Range
  • Working range: 3.0-2000.0 ng/mL
Precision Intra-lot Precision
  •  Intra-lot precision was determined by testing of AFP reference materials using 10 test devices from the same lot. CV ≤ 15%.
Inter-lot Precision
  • Inter-lot precision was determined by testing of AFP reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
Accuracy
  • AFP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
Linearity
  • A serial concentration of AFP controls at 3.0-2000.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
 

AMH FIA Test

Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    1. Range
    Working range: 0.1-25.0 ng/mL
    1. Precision
    Intra-lot Precision Intra-lot precision was determined by testing of AMH reference materials using 10 test devices from the same lot. CV ≤ 10%. Inter-lot Precision Inter-lot precision was determined by testing of AMH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
    1. Accuracy
    AMH control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
    1. Linearity
    A serial concentration of AMH reference materials at 0.1-25.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
      Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator

CEA FIA Test

Features:
      • Quantitative, non-subjective results
      • Increased sensitivity and broader detection range compared to rapid tests
      • Range
        •  Working range: 0.5-500.0 ng/mL
        Precision  Intra-lot Precision
        •  Intra-lot precision was determined by testing of CEA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
        Inter-lot Precision
        •  Inter-lot precision was determined by testing of CEA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
        Accuracy
        • CEA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        Linearity
        • A serial concentration of CEA reference materials at 1-500.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
        Hook Effect
        •  There is no Hook effect when the level is 100000.0 ng/mL.
        Reagents and Materials provided:
        1. Individually sealed foil pouches containing:
          1.  One test device
          2.  One desiccant
        2. Detection buffer tubes
        3. SD card
        4. Instructions for Use
        Materials required but not provided:
        1. Clock or timer
        2. RaFIA Immunofluorescence Analyzer
        3. RaFIA Immunofluorescence Incubator

CK-MB FIA Test

Features:
        • Quantitative, non-subjective results
        • Increased sensitivity and broader detection range compared to rapid tests
        RANGES
        • Working range: 1.0-80.0 ng/mL
        PRECISION
        • Intra-lot Precision Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV ≤ 15%.
        • Inter-lot Precision Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
        ACCURACY
        • CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        LINEARITY
        • A serial concentration of CK-MB reference materials at 2.5-80.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
        • REAGENTS AND MATERIALS PROVIDED:
          1. Individually sealed foil pouches containing: a. One test device b. One desiccant
          2. Detection buffer tubes
          3. SD card
          4. Instructions for Use
          MATERIALS REQUIRED BUT NOT PROVIDED:
          1. Clock or timer
          2. RaFIA Immunofluorescence Analyzer
          3. RaFIA Immunofluorescence Incubator

CRP FIA Test

Features:
    •  Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    RANGES
    • Working range: 2.5–200 mg/L
    PRECISION
    • Intra-lot Precision Intra-lot precision was determined by testing of CRP reference materials using 10 test devices from the same lot. CV ≤ 15%.
    • Inter-lot Precision Inter-lot precision was determined by testing of CRP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    ACCURACY
    • CRP control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
    LINEARITY
    • A serial concentration of CRP reference materials at 2.5–200 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
    • REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1.  Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0410

CRP FIA Test

25/kit

Serum, Plasma, Whole blood

cTnI FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 0.1–50.00 ng/mL
PRECISION: 
  • Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3002

cTnI FIA Test

25/kit

 Plasma, Whole blood

Cys-C FIA Test

Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    Range
    • Working range: 0.2-8.0 mg/L
    Precision Intra-lot Precision
    •  Intra-lot precision was determined by testing of Cys-C reference materials using 10 test devices from the same lot. CV ≤ 15%.
    Inter-lot Precision
    • Inter-lot precision was determined by testing of Cys-C reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    Accuracy
    • Cys-C control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
    Linearity
    • A serial concentration of Cys-C reference materials at 0.4-8.0 mg/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
      1. Individually sealed foil pouches containing:
        1. One test device
        2. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      Materials required but not provided:
      1. Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
 

D-Dimer FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–10.0 mg/L
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3010

D-Dimer FIA Test

25/kit

Serum, Plasma
 

Edan M16

Features:
  • Principle: GMR Technology
  • Innovative micro-array sensor and micro-fluidic technology
  • Sample Volume: 50μl
  • Display: 8.9 inch color LCD Display
  • Throughput: Results in 15 minutes after sample aspiration
  • Built-in thermal printer

Ferritin FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-1000.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of Ferritin reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of Ferritin reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • Ferritin control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of Ferritin reference materials at 2.0-1000.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F5010

Ferritin FIA Test

25/kit

Serum, Plasma
 

fPSA FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
Range
  • Working range: 0.1-30.0 ng/mL
Precision Intra-lot Precision
  • Intra-lot precision was determined by testing of fPSA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
Inter-lot Precision
  • Inter-lot precision was determined by testing of fPSA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
Accuracy
  • fPSA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
Linearity
  • A serial concentration of fPSA reference materials at 0.1-30.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator

FSH FIA Test

  Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    RANGES
    • Working range: 0.30-200.00 mIU/mL
    PRECISION
    • Intra-lot Precision Intra-lot precision was determined by testing of FSH reference materials using 10 test devices from the same lot. CV ≤ 15%.
    • Inter-lot Precision Inter-lot precision was determined by testing of FSH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    ACCURACY
    • FSH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%
    LINEARITY
    • A serial concentration of FSH reference materials at 1.00-200.00 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    • REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1. Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1100

FSH FIA Test

25/kit

Serum, Plasma