CTK Biotech

CEA FIA Test

Features:

      • Quantitative, non-subjective results
      • Increased sensitivity and broader detection range compared to rapid tests
      • Range
        •  Working range: 0.5-500.0 ng/mL

        Precision

         Intra-lot Precision

        •  Intra-lot precision was determined by testing of CEA reference materials using 10 test devices from the same lot.  CV ≤ 15%.

        Inter-lot Precision

        •  Inter-lot precision was determined by testing of CEA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

        Accuracy

        • CEA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.

        Linearity

        • A serial concentration of CEA reference materials at 1-500.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

        Hook Effect

        •  There is no Hook effect when the level is 100000.0 ng/mL.

        Reagents and Materials provided:

        1. Individually sealed foil pouches containing:
          1.  One test device
          2.  One desiccant
        2. Detection buffer tubes
        3. SD card
        4. Instructions for Use

        Materials required but not provided:

        1. Clock or timer
        2. RaFIA Immunofluorescence Analyzer
        3. RaFIA Immunofluorescence Incubator

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Category: Fluorescence Immunoassay Tests
  • Description
  • Additional information
Description

Features:

      • Quantitative, non-subjective results
      • Increased sensitivity and broader detection range compared to rapid tests
      • Range
        •  Working range: 0.5-500.0 ng/mL

        Precision

         Intra-lot Precision

        •  Intra-lot precision was determined by testing of CEA reference materials using 10 test devices from the same lot.  CV ≤ 15%.

        Inter-lot Precision

        •  Inter-lot precision was determined by testing of CEA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

        Accuracy

        • CEA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.

        Linearity

        • A serial concentration of CEA reference materials at 1-500.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

        Hook Effect

        •  There is no Hook effect when the level is 100000.0 ng/mL.

        Reagents and Materials provided:

        1. Individually sealed foil pouches containing:
          1.  One test device
          2.  One desiccant
        2. Detection buffer tubes
        3. SD card
        4. Instructions for Use

        Materials required but not provided:

        1. Clock or timer
        2. RaFIA Immunofluorescence Analyzer
        3. RaFIA Immunofluorescence Incubator
Additional information
Brand

CTK Biotech

Related products

Compare
Close

NT-proBNP FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 15–35000 pg/mL
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of NT-proBNP reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of NT-proBNP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • NT-proBNP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of NT-proBNP reference materials at 15–35000 pg/mL was tested, and the correlation coefficient (R) is ≥0.9900.
REAGENTS AND MATERIALS PROVIDED:
  1. Individually sealed foil pouches containing: a. One test device b. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

PRL FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 5.0-4000.0 μIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PRL reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of PRL reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PRL control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PRL reference materials at 5.0-4000.0 μIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1080

PRL FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

AMH FIA Test

Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    1. Range
    Working range: 0.1-25.0 ng/mL
    1. Precision
    Intra-lot Precision Intra-lot precision was determined by testing of AMH reference materials using 10 test devices from the same lot. CV ≤ 10%. Inter-lot Precision Inter-lot precision was determined by testing of AMH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
    1. Accuracy
    AMH control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
    1. Linearity
    A serial concentration of AMH reference materials at 0.1-25.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
      Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

β-hCG FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 5.0-200000.0 mIU/mL
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of β-hCG reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of β-hCG reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  •  β-hCG control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  •  A serial concentration of β-hCG reference materials at 5.0-200000.0 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
    4. REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1. Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1000

β-hCG FIA Test

25/kit

Serum, Plasma
Add to wishlist
Read more
Compare
Close

LH FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-200.0 mIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of LH reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of LH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • LH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of LH reference materials at 1.0-200.0 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1090

LH FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

cTnI FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 0.1–50.00 ng/mL
PRECISION: 
  • Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3002

cTnI FIA Test

25/kit

 Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

MAU FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 2.0-300.0 mg/L
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of MAU reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of MAU reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  • MAU control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of MAU reference materials at 2.0-300.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F2020

MAU FIA Test

25/kit

Urine
Add to wishlist
Read more
Compare
Close

MYO FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 2.5-400.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of MYO reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of MYO reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • MYO control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of MYO reference materials at 20.0-400.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3030

MYO FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Facebook