cTnI FIA Test

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests

RANGES:

  • Working range: 0.1–50.00 ng/mL

PRECISION: 

  • Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

ACCURACY:

  • cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.

LINEARITY:

  • A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    1. Individually sealed foil pouches containing:
      a. One test device
      b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use

    MATERIALS REQUIRED BUT NOT PROVIDED:

    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3002

cTnI FIA Test

25/kit

 Plasma, Whole blood

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Category: Fluorescence Immunoassay Tests
  • Description
  • Additional information
Description

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests

RANGES:

  • Working range: 0.1–50.00 ng/mL

PRECISION: 

  • Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

ACCURACY:

  • cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.

LINEARITY:

  • A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    1. Individually sealed foil pouches containing:
      a. One test device
      b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use

    MATERIALS REQUIRED BUT NOT PROVIDED:

    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3002

cTnI FIA Test

25/kit

 Plasma, Whole blood
Additional information
Brand

CTK Biotech

Related products

Compare
Close

CRP FIA Test

Features:
    •  Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    RANGES
    • Working range: 2.5–200 mg/L
    PRECISION
    • Intra-lot Precision Intra-lot precision was determined by testing of CRP reference materials using 10 test devices from the same lot. CV ≤ 15%.
    • Inter-lot Precision Inter-lot precision was determined by testing of CRP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    ACCURACY
    • CRP control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
    LINEARITY
    • A serial concentration of CRP reference materials at 2.5–200 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
    • REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1.  Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0410

CRP FIA Test

25/kit

Serum, Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

T4 FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 10.0-320.0 nmol/L
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of T4 reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of T4 reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
T4 control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of T4 reference materials at 20.0-320.0 nmol/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
The concentration of T4 is not greater than 15 nmol/L when the concentration of T3 is 500.0 ng/mL. The concentration of T4 is not greater than 15 nmol/L when the concentration of rT3 is 50.0 ng/mL. Product Specifications Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

tPSA FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.5-200.0 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of tPSA reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of tPSA reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
tPSA control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of tPSA reference materials at 0.5-200.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900. Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

D-Dimer FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–10.0 mg/L
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3010

D-Dimer FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

Total Vitamin D FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 5.0-120.0 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of Vitamin D reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of Vitamin D reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
Vitamin D control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of Vitamin D reference materials at 5.0-120.0ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900. Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. Solution A tube
  4. Solution B tube
  5. SD card
  6. Instructions for Use
  Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

PCT FIA Test

Features: PRODUCT FEATURES
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–100.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PCT reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of PCT reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PCT control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PCT reference materials at 0.1–100.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1.  Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0400

PCT FIA Test

25/kit

Serum, Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

CEA FIA Test

Features:
      • Quantitative, non-subjective results
      • Increased sensitivity and broader detection range compared to rapid tests
      • Range
        •  Working range: 0.5-500.0 ng/mL
        Precision  Intra-lot Precision
        •  Intra-lot precision was determined by testing of CEA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
        Inter-lot Precision
        •  Inter-lot precision was determined by testing of CEA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
        Accuracy
        • CEA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        Linearity
        • A serial concentration of CEA reference materials at 1-500.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
        Hook Effect
        •  There is no Hook effect when the level is 100000.0 ng/mL.
        Reagents and Materials provided:
        1. Individually sealed foil pouches containing:
          1.  One test device
          2.  One desiccant
        2. Detection buffer tubes
        3. SD card
        4. Instructions for Use
        Materials required but not provided:
        1. Clock or timer
        2. RaFIA Immunofluorescence Analyzer
        3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

LH FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-200.0 mIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of LH reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of LH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • LH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of LH reference materials at 1.0-200.0 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1090

LH FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Facebook