tPSA FIA Test

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range

Working range: 0.5-200.0 ng/mL

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of tPSA reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of tPSA reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

  1. Accuracy

tPSA control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of tPSA reference materials at 0.5-200.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Categories: Fluorescence Immunoassay Tests, Immunology
  • Description
  • Additional information
Description

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range

Working range: 0.5-200.0 ng/mL

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of tPSA reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of tPSA reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

  1. Accuracy

tPSA control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of tPSA reference materials at 0.5-200.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Additional information
Brand

CTK Biotech

Related products

Compare
Close

D-Dimer FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–10.0 mg/L
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3010

D-Dimer FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

PRL FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 5.0-4000.0 μIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PRL reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of PRL reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PRL control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PRL reference materials at 5.0-4000.0 μIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1080

PRL FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

T4 FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 10.0-320.0 nmol/L
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of T4 reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of T4 reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
T4 control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of T4 reference materials at 20.0-320.0 nmol/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
The concentration of T4 is not greater than 15 nmol/L when the concentration of T3 is 500.0 ng/mL. The concentration of T4 is not greater than 15 nmol/L when the concentration of rT3 is 50.0 ng/mL. Product Specifications Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

MYO FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 2.5-400.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of MYO reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of MYO reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • MYO control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of MYO reference materials at 20.0-400.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3030

MYO FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

Progesterone FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.30-80.00 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of PROG reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of PROG reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
PROG control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of PROG reference materials at 3.00-80.00 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
Crossed Factors Concentration Test Result
Estradiol 500 ng/mL < 0.3 ng/mL
Testosterone 500 ng/mL < 0.3 ng/mL
Cortisol 500 ng/mL < 0.3 ng/mL
Aldosterone 500 ng/mL < 0.3 ng/mL
Estriol 100 ng/mL < 0.3 ng/mL
17-β Estradiol 100 ng/mL < 0.3 ng/mL
Oestrone 100 ng/mL < 0.3 ng/mL
17-α hydroxyl progesterone 100 ng/mL < 0.3 ng/mL
Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

fPSA FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
Range
  • Working range: 0.1-30.0 ng/mL
Precision Intra-lot Precision
  • Intra-lot precision was determined by testing of fPSA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
Inter-lot Precision
  • Inter-lot precision was determined by testing of fPSA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
Accuracy
  • fPSA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
Linearity
  • A serial concentration of fPSA reference materials at 0.1-30.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

T3 FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.5–10.0 nmol/L
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of T3 reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of T3 reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
T3 control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of T3 reference materials at 0.8-10.0 nmol/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
The concentration of T3 is not greater than 1.0 nmol/L when the concentration of T4 is 500.0 ng/mL. The concentration of T3 is not greater than 1.0 nmol/L when the concentration of rT3 is 50.0 ng/mL. Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

AMH FIA Test

Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    1. Range
    Working range: 0.1-25.0 ng/mL
    1. Precision
    Intra-lot Precision Intra-lot precision was determined by testing of AMH reference materials using 10 test devices from the same lot. CV ≤ 10%. Inter-lot Precision Inter-lot precision was determined by testing of AMH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
    1. Accuracy
    AMH control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
    1. Linearity
    A serial concentration of AMH reference materials at 0.1-25.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
      Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Facebook