HomeImmunologyFluorescence Immunoassay Tests Total Vitamin D FIA Test
Previous product
Total IgE FIA Test
Back to products
Next product
tPSA FIA Test
Click to enlarge
CTK Biotech

Total Vitamin D FIA Test

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range

Working range: 5.0-120.0 ng/mL

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of Vitamin D reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of Vitamin D reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

  1. Accuracy

Vitamin D control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of Vitamin D reference materials at 5.0-120.0ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. Solution A tube
  4. Solution B tube
  5. SD card
  6. Instructions for Use

 

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Category: Fluorescence Immunoassay Tests
  • Description
  • Additional information
Description

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range

Working range: 5.0-120.0 ng/mL

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of Vitamin D reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of Vitamin D reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

  1. Accuracy

Vitamin D control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of Vitamin D reference materials at 5.0-120.0ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. Solution A tube
  4. Solution B tube
  5. SD card
  6. Instructions for Use

 

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Additional information
Brand

CTK Biotech

Related products

Compare
Close

FSH FIA Test

  Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    RANGES
    • Working range: 0.30-200.00 mIU/mL
    PRECISION
    • Intra-lot Precision Intra-lot precision was determined by testing of FSH reference materials using 10 test devices from the same lot. CV ≤ 15%.
    • Inter-lot Precision Inter-lot precision was determined by testing of FSH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    ACCURACY
    • FSH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%
    LINEARITY
    • A serial concentration of FSH reference materials at 1.00-200.00 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    • REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1. Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1100

FSH FIA Test

25/kit

Serum, Plasma
Add to wishlist
Read more
Compare
Close

fPSA FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
Range
  • Working range: 0.1-30.0 ng/mL
Precision Intra-lot Precision
  • Intra-lot precision was determined by testing of fPSA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
Inter-lot Precision
  • Inter-lot precision was determined by testing of fPSA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
Accuracy
  • fPSA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
Linearity
  • A serial concentration of fPSA reference materials at 0.1-30.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

AFP FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
Range
  • Working range: 3.0-2000.0 ng/mL
Precision Intra-lot Precision
  •  Intra-lot precision was determined by testing of AFP reference materials using 10 test devices from the same lot. CV ≤ 15%.
Inter-lot Precision
  • Inter-lot precision was determined by testing of AFP reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
Accuracy
  • AFP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
Linearity
  • A serial concentration of AFP controls at 3.0-2000.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
 
Add to wishlist
Read more
Compare
Close

TSH FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.4–100.0 μIU/mL
PRECISION
  • Intra-lot Precision: Intra-lot precision was determined by testing of TSH reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of TSH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤20%.
ACCURACY 
  • TSH control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of TSH reference materials at 0.4-100.0 μIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
REAGENTS AND MATERIALS PROVIDED:
  1. Individually sealed foil pouches containing: a. One test device b. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1030

TSH FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

Ferritin FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-1000.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of Ferritin reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of Ferritin reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • Ferritin control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of Ferritin reference materials at 2.0-1000.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F5010

Ferritin FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

MAU FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 2.0-300.0 mg/L
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of MAU reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of MAU reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  • MAU control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of MAU reference materials at 2.0-300.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F2020

MAU FIA Test

25/kit

Urine
Add to wishlist
Read more
Compare
Close

Total IgE FIA Test

Features: Quantitative, non-subjective results • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 1.0-1000.0 IU/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of IgE reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of IgE reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
  1. Accuracy
IgE control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of IgE reference materials at 1.0-1000 IU/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
No cross reactivities were observed while testing following immunoglobulins with indicated concentrations:
Immunoglobulins Concentrations Test Results
IgG 200 mg/mL ≤ 1.0 IU/mL
IgA 20 mg/mL ≤ 1.0 IU/mL
IgM 20 mg/mL ≤ 1.0 IU mL
Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
  Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

MYO FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 2.5-400.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of MYO reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of MYO reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • MYO control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of MYO reference materials at 20.0-400.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3030

MYO FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.