Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu

Manila:Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City

Davao: Gov. Generoso St.,Osmeña, Davao City

Lab Solutions Lab Solutions
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Lab Solutions Lab Solutions
Menu
HomeImmunologyFluorescence Immunoassay Tests MYO FIA Test
Previous product
MAU FIA Test
Back to products
Next product
NT-proBNP FIA Test
Click to enlarge
CTK Biotech

MYO FIA Test

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests

RANGES

  • Working range: 2.5-400.0 ng/mL

PRECISION

  • Intra-lot Precision
    Intra-lot precision was determined by testing of MYO reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision
    Inter-lot precision was determined by testing of MYO reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

ACCURACY

  • MYO control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.

LINEARITY

  • A serial concentration of MYO reference materials at 20.0-400.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing:
      a. One test device
      b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use

    MATERIALS REQUIRED BUT NOT PROVIDED:

    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3030

MYO FIA Test

25/kit

Serum, Plasma

 

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Categories: Fluorescence Immunoassay Tests, Immunology
  • Description
  • Additional information
Description

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests

RANGES

  • Working range: 2.5-400.0 ng/mL

PRECISION

  • Intra-lot Precision
    Intra-lot precision was determined by testing of MYO reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision
    Inter-lot precision was determined by testing of MYO reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

ACCURACY

  • MYO control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.

LINEARITY

  • A serial concentration of MYO reference materials at 20.0-400.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing:
      a. One test device
      b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use

    MATERIALS REQUIRED BUT NOT PROVIDED:

    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3030

MYO FIA Test

25/kit

Serum, Plasma
Additional information
Brand

CTK Biotech

Related products

Compare
Close

LH FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-200.0 mIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of LH reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of LH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • LH control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of LH reference materials at 1.0-200.0 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1090

LH FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

CK-MB FIA Test

Features:
        • Quantitative, non-subjective results
        • Increased sensitivity and broader detection range compared to rapid tests
        RANGES
        • Working range: 1.0-80.0 ng/mL
        PRECISION
        • Intra-lot Precision Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV ≤ 15%.
        • Inter-lot Precision Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
        ACCURACY
        • CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        LINEARITY
        • A serial concentration of CK-MB reference materials at 2.5-80.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
        • REAGENTS AND MATERIALS PROVIDED:
          1. Individually sealed foil pouches containing: a. One test device b. One desiccant
          2. Detection buffer tubes
          3. SD card
          4. Instructions for Use
          MATERIALS REQUIRED BUT NOT PROVIDED:
          1. Clock or timer
          2. RaFIA Immunofluorescence Analyzer
          3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

β-hCG FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 5.0-200000.0 mIU/mL
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of β-hCG reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of β-hCG reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  •  β-hCG control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  •  A serial concentration of β-hCG reference materials at 5.0-200000.0 mIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
    4. REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1. Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1000

β-hCG FIA Test

25/kit

Serum, Plasma
Add to wishlist
Read more
Compare
Close

Edan M16

Features:
  • Principle: GMR Technology
  • Innovative micro-array sensor and micro-fluidic technology
  • Sample Volume: 50μl
  • Display: 8.9 inch color LCD Display
  • Throughput: Results in 15 minutes after sample aspiration
  • Built-in thermal printer
Add to wishlist
Read more
Compare
Close

CRP FIA Test

Features:
    •  Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    RANGES
    • Working range: 2.5–200 mg/L
    PRECISION
    • Intra-lot Precision Intra-lot precision was determined by testing of CRP reference materials using 10 test devices from the same lot. CV ≤ 15%.
    • Inter-lot Precision Inter-lot precision was determined by testing of CRP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    ACCURACY
    • CRP control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
    LINEARITY
    • A serial concentration of CRP reference materials at 2.5–200 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
    • REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1.  Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0410

CRP FIA Test

25/kit

Serum, Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

MAU FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 2.0-300.0 mg/L
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of MAU reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of MAU reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  • MAU control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of MAU reference materials at 2.0-300.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F2020

MAU FIA Test

25/kit

Urine
Add to wishlist
Read more
Compare
Close

cTnI FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 0.1–50.00 ng/mL
PRECISION: 
  • Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3002

cTnI FIA Test

25/kit

 Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

Ferritin FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 1.0-1000.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of Ferritin reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of Ferritin reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • Ferritin control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of Ferritin reference materials at 2.0-1000.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F5010

Ferritin FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Facebook