T4 FIA Test

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range

Working range: 10.0-320.0 nmol/L

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of T4 reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of T4 reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

  1. Accuracy

T4 control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of T4 reference materials at 20.0-320.0 nmol/L was tested, and the correlation coefficient (R) is ≥ 0.9900.

  1. Specificity

The concentration of T4 is not greater than 15 nmol/L when the concentration of T3 is 500.0 ng/mL.

The concentration of T4 is not greater than 15 nmol/L when the concentration of rT3 is 50.0 ng/mL.

Product Specifications

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Category: Fluorescence Immunoassay Tests
  • Description
  • Additional information
Description

Features:

  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range

Working range: 10.0-320.0 nmol/L

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of T4 reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of T4 reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.

  1. Accuracy

T4 control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of T4 reference materials at 20.0-320.0 nmol/L was tested, and the correlation coefficient (R) is ≥ 0.9900.

  1. Specificity

The concentration of T4 is not greater than 15 nmol/L when the concentration of T3 is 500.0 ng/mL.

The concentration of T4 is not greater than 15 nmol/L when the concentration of rT3 is 50.0 ng/mL.

Product Specifications

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Additional information
Brand

CTK Biotech

Related products

Compare
Close

fPSA FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
Range
  • Working range: 0.1-30.0 ng/mL
Precision Intra-lot Precision
  • Intra-lot precision was determined by testing of fPSA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
Inter-lot Precision
  • Inter-lot precision was determined by testing of fPSA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
Accuracy
  • fPSA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
Linearity
  • A serial concentration of fPSA reference materials at 0.1-30.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

CK-MB FIA Test

Features:
        • Quantitative, non-subjective results
        • Increased sensitivity and broader detection range compared to rapid tests
        RANGES
        • Working range: 1.0-80.0 ng/mL
        PRECISION
        • Intra-lot Precision Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV ≤ 15%.
        • Inter-lot Precision Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
        ACCURACY
        • CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        LINEARITY
        • A serial concentration of CK-MB reference materials at 2.5-80.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
        • REAGENTS AND MATERIALS PROVIDED:
          1. Individually sealed foil pouches containing: a. One test device b. One desiccant
          2. Detection buffer tubes
          3. SD card
          4. Instructions for Use
          MATERIALS REQUIRED BUT NOT PROVIDED:
          1. Clock or timer
          2. RaFIA Immunofluorescence Analyzer
          3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

Total Vitamin D FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 5.0-120.0 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of Vitamin D reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of Vitamin D reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
Vitamin D control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of Vitamin D reference materials at 5.0-120.0ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900. Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. Solution A tube
  4. Solution B tube
  5. SD card
  6. Instructions for Use
  Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

HbA1c FIA Test

  Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 4%–15%
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of HbA1c reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of HbA1c reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • HbA1c control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of HbA1c reference materials at 4%–15% was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F6100

HbA1c FIA Test

25/kit

Serum, Plasma
Add to wishlist
Read more
Compare
Close

MAU FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 2.0-300.0 mg/L
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of MAU reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of MAU reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY:
  • MAU control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of MAU reference materials at 2.0-300.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F2020

MAU FIA Test

25/kit

Urine
Add to wishlist
Read more
Compare
Close

Testosterone FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.2-20.0 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of TES reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of TES reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
TES control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of TES reference materials at 0.2-20.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
Crossed Factors Concentration Test Result
Progesterone 500 ng/mL < 0.2 ng/mL
Estradiol 100 ng/mL < 0.2 ng/mL
Cortisol 500 ng/mL < 0.2 ng/mL
Aldosterone 500 ng/mL < 0.2 ng/mL
Estriol 100 ng/mL < 0.2 ng/mL
17α-Estradiol 500 ng/mL < 0.2 ng/mL
Oestrone 50 ng/mL < 0.2 ng/mL
Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    • One test device
    • One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

CEA FIA Test

Features:
      • Quantitative, non-subjective results
      • Increased sensitivity and broader detection range compared to rapid tests
      • Range
        •  Working range: 0.5-500.0 ng/mL
        Precision  Intra-lot Precision
        •  Intra-lot precision was determined by testing of CEA reference materials using 10 test devices from the same lot.  CV ≤ 15%.
        Inter-lot Precision
        •  Inter-lot precision was determined by testing of CEA reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
        Accuracy
        • CEA control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        Linearity
        • A serial concentration of CEA reference materials at 1-500.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
        Hook Effect
        •  There is no Hook effect when the level is 100000.0 ng/mL.
        Reagents and Materials provided:
        1. Individually sealed foil pouches containing:
          1.  One test device
          2.  One desiccant
        2. Detection buffer tubes
        3. SD card
        4. Instructions for Use
        Materials required but not provided:
        1. Clock or timer
        2. RaFIA Immunofluorescence Analyzer
        3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

PCT FIA Test

Features: PRODUCT FEATURES
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–100.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PCT reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of PCT reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PCT control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PCT reference materials at 0.1–100.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1.  Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0400

PCT FIA Test

25/kit

Serum, Plasma, Whole blood
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Facebook