CTK Biotech

Total IgE FIA Test

Features:

Quantitative, non-subjective results
• Increased sensitivity and broader detection range compared to rapid tests

  1. Range

Working range: 1.0-1000.0 IU/mL

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of IgE reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of IgE reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

  1. Accuracy

IgE control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of IgE reference materials at 1.0-1000 IU/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

  1. Specificity

No cross reactivities were observed while testing following immunoglobulins with indicated concentrations:

Immunoglobulins Concentrations Test Results
IgG 200 mg/mL ≤ 1.0 IU/mL
IgA 20 mg/mL ≤ 1.0 IU/mL
IgM 20 mg/mL ≤ 1.0 IU mL

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

 

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

Inquire about this Product

Product Inquiry

Compare
Add to wishlist
Categories: Fluorescence Immunoassay Tests, Immunology
  • Description
  • Additional information
Description

Features:

Quantitative, non-subjective results
• Increased sensitivity and broader detection range compared to rapid tests

  1. Range

Working range: 1.0-1000.0 IU/mL

  1. Precision

Intra-lot Precision

Intra-lot precision was determined by testing of IgE reference materials using 10 test devices from the same lot. CV ≤ 15%.

Inter-lot Precision

Inter-lot precision was determined by testing of IgE reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.

  1. Accuracy

IgE control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.

  1. Linearity

A serial concentration of IgE reference materials at 1.0-1000 IU/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.

  1. Specificity

No cross reactivities were observed while testing following immunoglobulins with indicated concentrations:

Immunoglobulins Concentrations Test Results
IgG 200 mg/mL ≤ 1.0 IU/mL
IgA 20 mg/mL ≤ 1.0 IU/mL
IgM 20 mg/mL ≤ 1.0 IU mL

Reagents and Materials provided:

  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use

 

Materials required but not provided:

  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Additional information
Brand

CTK Biotech

Related products

Compare
Close

AMH FIA Test

Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    1. Range
    Working range: 0.1-25.0 ng/mL
    1. Precision
    Intra-lot Precision Intra-lot precision was determined by testing of AMH reference materials using 10 test devices from the same lot. CV ≤ 10%. Inter-lot Precision Inter-lot precision was determined by testing of AMH reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
    1. Accuracy
    AMH control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
    1. Linearity
    A serial concentration of AMH reference materials at 0.1-25.0 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  • Reagents and Materials provided:
    1. Individually sealed foil pouches containing:
      1. One test device
      2. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
      Materials required but not provided:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

NT-proBNP FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 15–35000 pg/mL
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of NT-proBNP reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of NT-proBNP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • NT-proBNP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of NT-proBNP reference materials at 15–35000 pg/mL was tested, and the correlation coefficient (R) is ≥0.9900.
REAGENTS AND MATERIALS PROVIDED:
  1. Individually sealed foil pouches containing: a. One test device b. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

D-Dimer FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–10.0 mg/L
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of D-Dimer reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of D-Dimer reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY
  • D-Dimer control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of D-Dimer reference materials at 0.1–10.0 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3010

D-Dimer FIA Test

25/kit

Serum, Plasma
 
Add to wishlist
Read more
Compare
Close

Cys-C FIA Test

Features:
    • Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    Range
    • Working range: 0.2-8.0 mg/L
    Precision Intra-lot Precision
    •  Intra-lot precision was determined by testing of Cys-C reference materials using 10 test devices from the same lot. CV ≤ 15%.
    Inter-lot Precision
    • Inter-lot precision was determined by testing of Cys-C reference materials using 30 test devices from three consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    Accuracy
    • Cys-C control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
    Linearity
    • A serial concentration of Cys-C reference materials at 0.4-8.0 mg/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
      1. Individually sealed foil pouches containing:
        1. One test device
        2. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      Materials required but not provided:
      1. Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
 
Add to wishlist
Read more
Compare
Close

T3 FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.5–10.0 nmol/L
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of T3 reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of T3 reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
T3 control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of T3 reference materials at 0.8-10.0 nmol/L was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
The concentration of T3 is not greater than 1.0 nmol/L when the concentration of T4 is 500.0 ng/mL. The concentration of T3 is not greater than 1.0 nmol/L when the concentration of rT3 is 50.0 ng/mL. Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
Compare
Close

CRP FIA Test

Features:
    •  Quantitative, non-subjective results
    • Increased sensitivity and broader detection range compared to rapid tests
    RANGES
    • Working range: 2.5–200 mg/L
    PRECISION
    • Intra-lot Precision Intra-lot precision was determined by testing of CRP reference materials using 10 test devices from the same lot. CV ≤ 15%.
    • Inter-lot Precision Inter-lot precision was determined by testing of CRP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
    ACCURACY
    • CRP control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
    LINEARITY
    • A serial concentration of CRP reference materials at 2.5–200 mg/L was tested, and the correlation coefficient (R) is ≥0.9900.
    • REAGENTS AND MATERIALS PROVIDED:
      1. Individually sealed foil pouches containing: a. One test device b. One desiccant
      2. Detection buffer tubes
      3. SD card
      4. Instructions for Use
      MATERIALS REQUIRED BUT NOT PROVIDED:
      1.  Clock or timer
      2. RaFIA Immunofluorescence Analyzer
      3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0410

CRP FIA Test

25/kit

Serum, Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

cTnI FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 0.1–50.00 ng/mL
PRECISION: 
  • Intra-lot Precision: Intra-lot precision was determined by testing of cTnI reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of cTnI reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • cTnI control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of cTnI reference materials at 0.1–50.00 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F3002

cTnI FIA Test

25/kit

 Plasma, Whole blood
Add to wishlist
Read more
Compare
Close

CK-MB FIA Test

Features:
        • Quantitative, non-subjective results
        • Increased sensitivity and broader detection range compared to rapid tests
        RANGES
        • Working range: 1.0-80.0 ng/mL
        PRECISION
        • Intra-lot Precision Intra-lot precision was determined by testing of CK-MB reference materials using 10 test devices from the same lot. CV ≤ 15%.
        • Inter-lot Precision Inter-lot precision was determined by testing of CK-MB reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
        ACCURACY
        • CK-MB control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
        LINEARITY
        • A serial concentration of CK-MB reference materials at 2.5-80.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
        • REAGENTS AND MATERIALS PROVIDED:
          1. Individually sealed foil pouches containing: a. One test device b. One desiccant
          2. Detection buffer tubes
          3. SD card
          4. Instructions for Use
          MATERIALS REQUIRED BUT NOT PROVIDED:
          1. Clock or timer
          2. RaFIA Immunofluorescence Analyzer
          3. RaFIA Immunofluorescence Incubator
Add to wishlist
Read more
About LabSolution Technologies

LabSolution Technologies, Inc. has its headquarters in Cebu City, “The Queen City of the South”. A professional healthcare company established on March 2006 as ALP Diagnostix and later incorporated as LabSolution Technologies, Inc. on June 2008.

READ MORE

Locations

  • Cebu: 3rd floor, ALP Tower, 609 Tres de Abril St, Cebu City, Cebu
  • Manila: Unit A, # 286 El Grande Avenue, Phase 3, BF Homes, Parañaque City
  • Davao: Gov. Generoso St., Osmeña, Davao City

Latest Products
  • Chemistry Reagents
  • DiaSys Respons 940
  • DiaSys Respons 920
2023. Lab Solutions. All Rights Reserved.
  • Home
  • Products
  • Services
  • About us
  • Blog
  • Contact us
Facebook