NT-proBNP FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES:
  • Working range: 15–35000 pg/mL
PRECISION:
  • Intra-lot Precision: Intra-lot precision was determined by testing of NT-proBNP reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision: Inter-lot precision was determined by testing of NT-proBNP reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 15%.
ACCURACY:
  • NT-proBNP control materials with two different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY:
  • A serial concentration of NT-proBNP reference materials at 15–35000 pg/mL was tested, and the correlation coefficient (R) is ≥0.9900.
REAGENTS AND MATERIALS PROVIDED:
  1. Individually sealed foil pouches containing: a. One test device b. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
MATERIALS REQUIRED BUT NOT PROVIDED:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

PCT FIA Test

Features: PRODUCT FEATURES
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 0.1–100.0 ng/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PCT reference materials using 10 test devices from the same lot. CV ≤ 15%.
  • Inter-lot Precision Inter-lot precision was determined by testing of PCT reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PCT control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PCT reference materials at 0.1–100.0 ng/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1.  Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F0400

PCT FIA Test

25/kit

Serum, Plasma, Whole blood

PRL FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
RANGES
  • Working range: 5.0-4000.0 μIU/mL
PRECISION
  • Intra-lot Precision Intra-lot precision was determined by testing of PRL reference materials using 10 test devices from the same lot. CV ≤ 15%
  • Inter-lot Precision Inter-lot precision was determined by testing of PRL reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
ACCURACY
  • PRL control materials with three different concentrations were tested by every lot of test device, and the deviations were within ±15%.
LINEARITY
  • A serial concentration of PRL reference materials at 5.0-4000.0 μIU/mL was tested, and the correlation coefficient (R) is ≥0.9900.
  • REAGENTS AND MATERIALS PROVIDED:
    1. Individually sealed foil pouches containing: a. One test device b. One desiccant
    2. Detection buffer tubes
    3. SD card
    4. Instructions for Use
    MATERIALS REQUIRED BUT NOT PROVIDED:
    1. Clock or timer
    2. RaFIA Immunofluorescence Analyzer
    3. RaFIA Immunofluorescence Incubator
CATALOG # PRODUCT QUANTITY SPECIMEN

F1080

PRL FIA Test

25/kit

Serum, Plasma
 

Progesterone FIA Test

Features:
  • Quantitative, non-subjective results
  • Increased sensitivity and broader detection range compared to rapid tests
  1. Range
Working range: 0.30-80.00 ng/mL
  1. Precision
Intra-lot Precision Intra-lot precision was determined by testing of PROG reference materials using 10 test devices from the same lot. CV ≤ 15%. Inter-lot Precision Inter-lot precision was determined by testing of PROG reference materials using 30 test devices from 3 consecutive lots randomly (10 test devices from each lot). CV ≤ 20%.
  1. Accuracy
PROG control materials with two different concentrations were tested by every lot of test device, and the deviations were within ± 15%.
  1. Linearity
A serial concentration of PROG reference materials at 3.00-80.00 ng/mL was tested, and the correlation coefficient (R) is ≥ 0.9900.
  1. Specificity
Crossed Factors Concentration Test Result
Estradiol 500 ng/mL < 0.3 ng/mL
Testosterone 500 ng/mL < 0.3 ng/mL
Cortisol 500 ng/mL < 0.3 ng/mL
Aldosterone 500 ng/mL < 0.3 ng/mL
Estriol 100 ng/mL < 0.3 ng/mL
17-β Estradiol 100 ng/mL < 0.3 ng/mL
Oestrone 100 ng/mL < 0.3 ng/mL
17-α hydroxyl progesterone 100 ng/mL < 0.3 ng/mL
Reagents and Materials provided:
  1. Individually sealed foil pouches containing:
    1. One test device
    2. One desiccant
  2. Detection buffer tubes
  3. SD card
  4. Instructions for Use
Materials required but not provided:
  1. Clock or timer
  2. RaFIA Immunofluorescence Analyzer
  3. RaFIA Immunofluorescence Incubator

PSA Semi-quantitative Rapid Test

Features:
  • Non-invasive screening method for prostate cancer
  • Three-line test result to facilitate determination of patient’s PSA level
  • A cut-off value of 4 ng/mL and a reference value of 10 ng/mL for easy interpretation of the semi-quantitative test result
  • Individually sealed foil pouches containing:
    • One cassette device
    • One desiccant
  • Plastic droppers
  • Sample diluent (REF SB-R2002, 5 mL/bottle)
  • One package insert (instruction for use)
CATALOG # PRODUCT QUANTOTY SPECIMEN

R2002C

PSA Semi-Quantitative Rapid Test

30/kit

S, P & WB

Rayto Chemray 120

Features:
  • Automatic, Random Access
  • Throughput: 120 tests per hour
  • Reagent cooling and pre-heating, liquid level detection
  • Analytical modes: End-point, two-point, kinetic
  • Sample Position: 9 positions
  • Reagent Position: 26 positions
  • Refrigerated reagent tray
  • Reaction Position: 90 cuvettes, 10 cuvettes per strip
  • Reaction Volume: 180-500μl
  • Probe: collision protection, Liquid Level Detection and inventory checking

Rayto Chemray 330

Features:
  • Automatic, Random Access
  • Analytical Modes: End-point, two-point, kinetic
  • Throughput: Constant speed 200 tests per hour
  • Automatic 8-step washing system, low carry-over
  • 24 hours non-stop reagent cooling compartment
  • Sample Position: 40 positions
  • Reagent Position: 40 reagent positions in refrigerated compartment
  • Reaction Position: 81 cuvettes, 9 cuvette per strip
  • Probe: Collision protection, Liquid level detection and inventory checking

Rayto Chemray-240

Features:
  • Automatic, Random Access
  • Throughput: up to 160 tests per hour
  • Analytical Modes: End-point, Two-point, Kinetic
  • Automatic 8-steps washing system
  • Sample Position: 40 positions
  • Reagent Position: 40 positions in refrigerated compartment
  • Reaction Position: 81 cuvettes, 9 cuvettes per strip
  • Reaction Volume: 180-500μl
  • Independent mixing probe

Rayto RT-9200

Features: • Easy keypad Operation with LCD display • Wavelength from 330-800nm; 5 standard filters, 3 optional • Supports flowcell and cuvette mode • Analytical modes include kinetic, two-point, bichromatic end-point with or without reagent blank or sample blank, linear or non-linear calibration curve • Memory of 60 programs and 2200 test results • Excellent QC function • Internal thermal-sensitive printer  

RIELE Photometer 4040

Features:
  • Type: Single-beam filter photometer
  • Light Source: LED with  long lifetime
  • Wavelength: 330-730nm; 6 standard filters + 3 optional filter of choice
  • Cuvette System: Normal standard cuvette (macro or semi-micro, disposable or special optical glass)
  • Memory: capacity for up to 231 programmable methods
  • Data Logging: up to 1000 test results can be saved
  • Quality Control: up to 50 methods, Levey-Jennings plot
  • Integrated Printer: Thermal printer, 24 characters per line

RIELE Photometer 5010

Features:
  • Type: Semi-automatic, single-beam filter photometer
  • Light Source: LED with long lifetime
  • Wavelength: 340 to 730nm; 6 standard filters + 3 optional filter of choice
  • Cuvette System: Micro flow cell; 32μl, 10 mm light path interchangeable with normal standard cuvettes (macro or semi-micro, disposable or special optical glass)
  • Aspiration System: Built-in peristaltic pump driven by stepper motor
  • Memory: Capacity for up to 231 programmable methods
  • Data Logging: up to 1000 test results can be saved
  • Quality Control: up to 50 methods, Levey-Jennings plot
  • Integrated Printer: Thermal printer, 24 characters per line

Rota/Adeno Ag Rapid Test CE

Features:
  • Stool collection device minimizes sample handling
  • Time saving differentiation between rotavrus and adenovirus in a single test
  • Result is available in 15 minutes
  • No cross-reactivity with other gastrointestinal infectious diseases
  • Individually sealed foil pouches containing:
    • One cassette device
    • One desiccant
  • Stool collection devices, each containing 2 mL extraction buffer
  • Plastic droppers (for transferring watery stool)
  • Patient ID stickers
  • Package insert (instruction for use)
CATALOG # PRODUCT QUANTOTY SPECIMEN

R0196C

 Rota/Adeno Ag Rapid Test CE

25/kit

Fecal